Cerclage Parahisian Septal Pacing through the Septal Perforator Branch of the Great Cardiac Vein: Bedside-to-Bench Development of a Novel Technique and Lead

Min Soo Cho et al.

Heart Rhythm · 2019

Summary

Background

Cerclage parahisian septal pacing places a pacing lead in the interventricular septum near the His bundle through the septal perforator branch of the great cardiac vein, the catheter route first used for transcatheter mitral cerclage annuloplasty.

Methods

In a prospective study (NCT03438591, 2016–2018), five patients underwent cerclage pacing with commercial coronary vein leads. Based on this experience, a purpose-built quadripolar cerclage pacing lead (2.4F–4F, monorail design, 3-mm electrode spacing) was developed and tested in seven swine.

Results

Lead implantation and pacing succeeded in all five patients, with intraseptal pacing in four. Procedures were long (median 227 minutes; fluoroscopy 57 minutes), one lead dislodged on the day after the procedure, and the clinical protocol was suspended. In swine, narrower electrode spacing gave paced QRS durations closer to native (paced/native ratio 1.05 ± 0.11 with 10-mm vs 1.37 ± 0.17 with 20-mm spacing), and the purpose-built lead crossed the septum without snare countertraction, with QRS duration similar to sinus rhythm.

Conclusions

The authors describe the initial feasibility of cerclage parahisian septal pacing and note that long-term durability of the lead and the risk of dislodgment must be addressed before clinical application.

No abstract is available for this article. This summary was prepared by Tau Medical from the full text; read the full article on the publisher's site.