Background
Current treatments for hypertrophic obstructive cardiomyopathy (HOCM), including medication, surgery, and alcohol septal ablation (ASA), have limitations in terms of efficacy and safety. To address these challenges, we developed the trans-coronary intra-septal radiofrequency ablation (TIRA) device.
Methods
This first-in-human trial was conducted to assess the efficacy and safety of the TIRA device. Moreover, evaluations were conducted before the procedure and at 3, 6, and 12 months post-procedure using computed tomography, magnetic resonance imaging, echocardiography, and the 6 min walk distance (6MWD) test.
Results
Four patients were enrolled, and follow-up imaging at 3, 6, and 12 months showed a reduction in the interventricular septal (IVS) thickness (baseline mean: 22.6 mm; 12-month mean: 18.9 mm) and a decrease in the LVOT pressure gradient at 12 months (resting baseline mean: 84.64 mmHg; resting 12-month mean: 43.56 mmHg; Valsalva baseline mean: 129.96 mmHg; Valsalva 12-month mean: 108.16 mmHg). However, reductions in the IVS thickness on echocardiography and improvements in 6MWD were observed in only two patients.
Conclusions
No significant adverse events, such as arrhythmias or vascular injuries, were reported. These findings suggest that the TIRA device may be a safe and effective option for treating HOCM. However, further studies are required to confirm these results.
Abstract as published in Biomedicines, 2024;12(12), reproduced without changes (source: PubMed PMID 39767669). Read the full article on the publisher's site.
Related publications
Bioengineering 2025 · Preclinical
Gradual Temperature Rise in Radiofrequency Ablation: Enhancing Lesion Quality and Safety in Porcine Myocardial Tissue
View detailsJACC: BTS 2020 · Preclinical