논문 목록은 원문 그대로 영어로만 제공됩니다.한국어 홈으로 돌아가기

Clinical Evidence / Publications / Clinical

  • Clinical
  • Tricuspid

Safety and Efficacy of Pivot-Bridge for Severe Functional Tricuspid Regurgitation: The First-in-Human Experiences

Eun Kyoung Kim et al.

Catheterization and Cardiovascular Interventions · 2026

Summary

Fifteen patients with severe functional TR were treated under fluoroscopy and TTE guidance. Procedural success was achieved in all patients, TR was reduced by at least one grade in every patient, and no serious adverse events occurred during implantation.

Part of the research behind the Pivot‑TR platform for tricuspid regurgitation.

Key results

100%

procedural success

−50%

average vena contracta

0

serious adverse events during implantation

2.5 → 1.5

average NYHA class

Study at a glance

Design
Multicenter, single-arm, short-term implantation study
Sites
11 hospitals in Korea
Patients
15 enrolled of 19 screened
Implant duration
Up to 7 days (temporary)
Imaging
Echocardiography and cardiac CT with core-lab review

Results from a single-arm study of 15 patients with temporary implantation; there was no control group. Pivot Bridge is an investigational device, and its safety and effectiveness have not been established.