Abstract
Background
Tricuspid regurgitation is treated by valve repair or replacement. However, these methods have limitations, and alternative treatment methods are therefore required.
Objectives
In this study, a new method of tricuspid valve treatment using artificial membrane insertion is analyzed. We performed tricuspid valve simulations using an artificial membrane inserted into the right ventricle (RV) or right atrium (RA).
Methods
We use the lattice Boltzmann method with the immersed boundary condition to model the structural motion of the valve leaflet. The effect of membrane insertion is analyzed in terms of the stress, force, and impulse on the valve leaflet, along with the velocity, pressure, jet volume, and Reynolds stress in the flow field.
Results
While the use of either membrane (RA or RV) leads to improved valve closure relative to the use of no membrane, the RV membrane is more effective than the RA membrane in achieving improved valve closure. In addition, a larger membrane area with a shorter distance between the leaflet and membrane increases membrane efficacy.
Conclusion
Our results suggest that membrane insertion can form an effective new method for the treatment of tricuspid regurgitation.
Abstract as published in Computer methods and programs in biomedicine, 2020;191:105421, reproduced without changes (source: PubMed PMID 32146209). Read the full article on the publisher's site.